Software Verification & Validation (IEC 62304)

MDR Reference: Annex II, Section 6.1.2; Annex I GSPRs 17.1-17.4
BPG Rev.4 Reference: Section 6.1.2
Service Item ID: mdr_sec06_software_vv
Applicability: Devices containing software, Software as a Medical Device (SaMD), AI/ML-based devices

Overview

Software V&V demonstrates that medical device software is developed, verified, and validated according to a structured lifecycle process appropriate to its safety classification. The primary standard is IEC 62304, supplemented by IEC 82304-1 for health software products and MDCG 2019-16 for cybersecurity (covered separately in Cybersecurity).

Current Standard: IEC 62304

Edition 1 (Current Harmonised Version)

EN 62304:2006+A1:2015 remains the harmonised version under EU MDR as of August 2026.

Edition 2 (Publication: August 2026)

IEC 62304 Edition 2 is scheduled for publication on 12 August 2026. Key changes:

  • Scope expansion: From "medical device software" to "health software" (broader scope)
  • Safety classification simplified: Three classes (A/B/C) replaced by two Software Process Rigor Levels (I and II)
  • AI/ML lifecycle: New clause 5.1.15 plus Annex E for AI/ML-specific requirements
  • Cybersecurity integration: Threat modelling, secure-by-design, vulnerability management as core design controls
  • ISO 13485 and ISO 14971: Removed as normative references (still applied in practice for regulated devices)
  • Agile guidance: New Annex F (informative), based on AAMI TIR45:2023

Transition note: Regulatory recognition of Edition 2 is expected 2-3 years after publication. EN 62304:2006+A1:2015 stays the state of the art for MDR compliance until Edition 2 is harmonised (estimated 2028-2029). New programs should consider designing to Edition 2 early.

Key Deliverables

DocumentDescription
Software Safety ClassificationClass A/B/C justification (Ed.1) or Rigor Level I/II (Ed.2)
Software Development PlanLifecycle model, standards, tools, V&V strategy
Software Requirements SpecificationFunctional, non-functional, GSPR 17.4, cybersecurity
Software Architecture DesignComponents, SOUP list, interfaces
Software Detailed DesignUnit-level design data (Class B & C / Rigor II)
SOUP ListName, version, manufacturer, functional requirements
V&V Plans, Protocols, ReportsUnit, integration, system testing
Traceability MatrixRequirements -> Architecture -> Tests -> Results
Software Release DocumentationRelease notes, known anomalies, residual risks

Safety Classification (IEC 62304 Ed.1)

ClassSoftware Contribution to Hazardous SituationProcess Rigor
ANo contribution to hazardous situation, or hazard controlled externallyMinimal
BCan contribute to hazardous situation that does NOT result in serious injuryModerate
CCan contribute to hazardous situation resulting in death or serious injuryFull

AI/ML-Specific Requirements

For AI/ML-based medical devices, additional documentation is expected:

  • Training and testing dataset documentation (provenance, size, quality)
  • Model architecture and hyperparameter selection rationale
  • Performance metrics and validation methodology
  • Monitoring triggers and retraining decision framework
  • EU Ethics Guidelines for Trustworthy AI compliance
  • Team-NB "AI in Medical Devices" questionnaire (if applicable)

See also: MDCG 2025-6: MDR/IVDR & AI Act

GSPR Mapping

GSPRRequirement
17.1Software lifecycle according to state of the art
17.2Software development and lifecycle processes
17.3Mobile platforms -- IT environment requirements
17.4Cybersecurity requirements (see Cybersecurity)