Measuring / Diagnostic Function Validation

MDR Reference: Annex I, GSPR 15; Classification Rule 10 (Class I(m))
Service Item ID: mdr_sec06_measuring
Applicability: Devices with a measuring function, diagnostic devices

Overview

Devices with a measuring or diagnostic function must demonstrate the accuracy, precision, and reliability of their measurement capabilities. MDR GSPR 15 establishes specific requirements, and Rule 10 may trigger Class I(m) classification for measuring devices.

Key Requirements

  • Accuracy and precision: Documented and validated against reference methods
  • Calibration: Traceable to national or international metrological standards
  • Measurement limits: Defined and validated (range, resolution, linearity)
  • Metrological traceability: Per ISO/IEC 17025 or equivalent
  • Clinical performance: Sensitivity, specificity, positive/negative predictive values (for diagnostics)

Key Deliverables

DocumentDescription
Measurement Performance SpecificationAccuracy, precision, range, resolution
Calibration Protocol & ReportMetrological traceability evidence
Analytical Performance ReportRepeatability, reproducibility, linearity, interference
Clinical Performance DataSensitivity, specificity (if applicable)
Labelling of Measurement AccuracyPer MDR Annex I GSPR 15 requirements

GSPR Mapping

GSPRRequirement
15.1Measuring devices designed for accuracy and stability
15.2Measurement values expressed in legal units
15.3Accuracy limits indicated
  • Design V&V -- Measurement validation as part of design V&V
  • Software V&V -- Software algorithms performing measurements