Periodic Safety Update Report (PSUR)
MDR Reference: Article 86; Annex III
BPG Rev.4 Reference: Section 7.2 (pp. 90-92)
Document Code: AIII-PSUR
Responsible Party: CRO
Service Mode: Write only
Applicability
- Required for: Class IIa, IIb, and III devices
- Update frequency (per Article 86):
- Class IIa: at least every 2 years
- Class IIb / Class III: at least annually
- Class I devices: PMS Report per Article 85 (not PSUR)
Requirements Checklist
Content (per Article 86 and Annex III)
- [ ] Summary of PMS data analysis for the reporting period
- [ ] Conclusions from benefit-risk assessment based on PMS data
- [ ] Main findings from PMCF activities
- [ ] Volume of sales data (units sold) and estimation of usage population
- [ ] Summary of complaints (categorised by type and frequency)
- [ ] Serious incidents and FSCAs reported during period
- [ ] Trend analysis against PMS Plan indicators and thresholds
- [ ] Actions taken as result of PMS findings (CAPAs, design changes, labelling updates)
- [ ] Conclusions and recommendations:
- [ ] Risk-benefit ratio remains acceptable
- [ ] Actions to optimise safety and performance
- [ ] Any necessary updates to TD (CER, RMR, IFU, GSPR, etc.)
- [ ] Reference to PMCF evaluation report findings
- [ ] Updates to SSCP triggered (if applicable)
Lifecycle Integration
- [ ] PSUR conclusions reflected in next CER update
- [ ] PSUR findings feed back into risk management file
- [ ] PSUR triggers PMS Plan updates where needed
- [ ] Alignment with SSCP annual review
- [ ] Alignment with NB surveillance schedule
Submission
- [ ] Uploaded to EUDAMED (when functional)
- [ ] Available to NB for assessment
- [ ] In absence of EUDAMED: follow NB-specific submission arrangements (per MDCG 2021-1)
Key Regulatory References
- MDR (EU) 2017/745, Article 86 (PSUR)
- MDR (EU) 2017/745, Annex III (PMS documentation)
- MDR (EU) 2017/745, Article 87 (Vigilance reporting)
- MDCG 2022-21: Guidance on PSUR (including end of obligation to update)
- MDCG 2021-1: EUDAMED interim arrangements
- Team-NB BPG for TD Submission, Rev.4 (2026-04-21), Section 7.2
Common Pitfalls (from BPG Rev.4)
- PSUR update frequency not per MDR Article 86 requirements
- Incomplete trend analysis or missing connection to PMS Plan indicators
- Benefits section is a re-statement of CER rather than analysis of new post-market data
- Actions taken not clearly linked to specific PMS findings
- Serious incident analysis lacking root cause investigation
- PSUR not aligned with PMCF report and SSCP updates
Dependency
PSUR is a Phase 5 (post-certification lifecycle) deliverable:
- Depends on PMS Plan (defines data collection scope)
- Incorporates PMCF Report findings
- Triggers updates to CER, RMR, IFU, SSCP
- Feeds into NB annual surveillance review