Skip to content

Product Verification & Validation (V&V)

MDR Reference: Annex II, Section 6 (pre-clinical evidence)
BPG Rev.4 Reference: Section 6 (pp. 32-62)
Document Code: AII-S6-VV

MDR Annex II Section 6 requires manufacturers to provide the results of all V&V tests, studies, and assessments that demonstrate conformity with the applicable GSPRs. This section covers pre-clinical evidence including bench testing, biocompatibility, software V&V, electrical safety, sterilisation, packaging, usability, and special pre-clinical studies.

V&V Sub-topics

TopicKey StandardsApplicability
Design V&V (Performance & Safety)ISO 13485, product-specific standardsAll devices
Biocompatibility EvaluationISO 10993-1:2025 (Ed.6)Devices with body contact
Electrical Safety & EMCEN 60601-1 (+A13:2024), EN 60601-1-2Active electrical devices
Software V&VIEC 62304:2006+A1:2015 (Ed.2 Aug 2026)Software / SaMD
CybersecurityEN IEC 81001-5-1:2022, MDCG 2019-16 Rev.1Connected software devices
Stability & Shelf-lifeASTM F1980, ASTM D4169Non-software devices with shelf-life claims
Packaging ValidationEN ISO 11607-1/-2Sterile / packaged devices
Sterilisation ValidationISO 11135, ISO 11137, ISO 17665Sterile devices
Usability / Human FactorsIEC 62366-1:2015+A1:2020All devices
Device Lifetime ValidationProduct-specific, MDR GSPR 6.6Non-software devices
Measuring / Diagnostic FunctionProduct-specific, MDR GSPR 15Measuring / diagnostic devices
MRI Safety (Implants)ASTM F2052/F2213/F2119/F2182, ISO/TS 10974Implantable devices
Device InteroperabilityMDR GSPR 14.5Active devices with interfaces
Reprocessing ValidationISO 17664Reusable devices
Drug-Device CombinationMDR Art.1(8)(9), CTD Module 3Drug-device combinations
Animal Tissue DevicesReg (EU) 722/2012, ISO 22442Animal-origin tissue devices
Human Tissue DevicesMDR GSPR 13.2Human-origin tissue devices
Absorbed / Dispersed SubstancesMDR Rule 21Devices with absorbed substances
CMR / Endocrine Disrupting SubstancesMDR GSPR 10.4.1Devices with CMR/ED substances

General V&V Principles

Per MDR Annex II Section 6.1:

  • All V&V results must be critically evaluated, not merely presented
  • Each test must be linked to specific GSPRs and/or risk control measures
  • For existing devices without new testing: rationale per Annex II 6.1(b) required
  • Protocols must include justified sample sizes, acceptance criteria, and confidence intervals
  • Worst-case and representative testing must be justified
  • Contract test laboratory accreditation/certification must be referenced
  • Deviations must be investigated and documented with acceptance rationale

Key Regulatory References

  • MDR (EU) 2017/745, Annex II, Section 6
  • Team-NB BPG for TD Submission, Rev.4 (2026-04-21), Section 6
  • MDCG Guidance Documents

Content licensed under CC BY 4.0