Skip to content

Design Verification & Validation (Performance & Safety)

MDR Reference: Annex II, Section 6.1 (pre-clinical data)
BPG Rev.4 Reference: Section 6.1.5
Service Item ID: mdr_sec06_design_vv
Applicability: All medical devices

Overview

Design V&V is the systematic process of confirming that device design outputs meet design inputs (verification) and that the finished device meets user needs and intended uses (validation). Under MDR, manufacturers must provide evidence that their device conforms to the applicable GSPRs through a structured design control process.

Scope

  • Design verification: confirms design outputs meet design input requirements (bench testing, analysis, inspection)
  • Design validation: confirms the finished device meets user needs under actual or simulated use conditions
  • Performance testing: demonstrates device meets its claimed performance specifications
  • Safety testing: demonstrates device does not introduce unacceptable risks under normal and fault conditions

Key Deliverables

DocumentDescription
Design Control MatrixTraceability from user needs -> design inputs -> design outputs -> V&V evidence
Verification Plan & ProtocolsTest methods, sample sizes, acceptance criteria for each design output
Verification ReportsTest results, pass/fail, deviations, statistical analysis
Validation Plan & ProtocolsClinical/simulated-use conditions, user population, acceptance criteria
Validation ReportsResults demonstrating intended use fulfilment

Regulatory Requirements

MDR Annex II Section 6.1 Requirements

  • Results and critical evaluation of all V&V tests/studies
  • Each test linked to specific GSPRs and/or risk control measures
  • All design requirements and specification documents provided
  • Design control input/output traceability matrix
  • Overarching design V&V plan(s) and associated report(s)
  • Protocols with justified sample sizes (standards-based or risk-based), acceptance criteria, confidence intervals
  • Worst-case and representative testing justification
  • Pre-conditioning documented in advance with data included in reports
  • Deviations/discrepancies investigated and documented with acceptance rationale
  • Contract test laboratory accreditation/certification referenced

GSPR Mapping

GSPRRequirementV&V Evidence
1Safe and perform as intendedOverall V&V conclusion
3State of the artCurrent standards applied
6.1Chemical, physical, biological propertiesMaterial and performance testing
8Device and manufacturing process designDesign control documentation

Best Practice Guidance

  • Follow ISO 13485:2016 Clause 7.3 (Design and Development) for design control structure
  • Use a Design History File (DHF) to compile all design control records
  • Statistical justification per ASTM E2709 or equivalent for sample sizes
  • For historic testing on legacy devices: clear identification of which reports apply to the current device version per Annex II 6.1(b)

Key Standards

StandardTitleStatus
ISO 13485:2016Medical devices -- QMS -- Requirements for regulatory purposesHarmonised (EN ISO 13485:2016+A11:2021)
ISO 14971:2019Application of risk management to medical devicesHarmonised (EN ISO 14971:2019+A11:2021)

Note: Product-specific performance standards vary by device type (e.g., EN ISO 21535 for hip implants, EN ISO 25539 for vascular devices). Consult the harmonised standards list and other applicable standards for your device category.

Content licensed under CC BY 4.0