Article 10 — General Obligations of Manufacturers ​

Source: EU MDR 2017/745 Article 10 | OJ L 117, 5.5.2017

Core Requirements ​

Before placing a device on the market or putting it into service, manufacturers shall ensure that the device has been designed and manufactured in accordance with this Regulation.

10.1 — Quality Management System (QMS) ​

Manufacturers shall establish, document, implement, maintain, update and continuously improve a quality management system covering at least:

  • A regulatory compliance strategy, including conformity assessment procedures and device change management
  • Identification of applicable GSPRs and exploration of options to meet those requirements
  • Responsibilities for design and development of devices
  • Management of production and service provision
  • Risk management (see Annex I GSPR)
  • Clinical evaluation (see Annex XIV), including PMCF
  • Labelling and packaging
  • Post-market surveillance (PMS), including PMCF (see Annex III)
  • Communication with competent authorities, notified bodies, economic operators, customers and/or other relevant parties
  • Processes for information transfer and coordination between economic operators
  • Reporting of serious incidents and field safety corrective actions
  • Corrective and preventive actions and verification of their effectiveness

10.2 — Technical Documentation ​

Manufacturers shall draw up, keep up to date and make available technical documentation in accordance with Annex II and Annex III.

10.3 — Conformity Assessment ​

Manufacturers shall subject their devices to the applicable conformity assessment procedure as set out in Article 52.

10.4 — EU Declaration of Conformity ​

Manufacturers shall draw up an EU declaration of conformity in accordance with Article 19 and affix the CE marking to the device (see Article 20).

10.9 — Authorised Representative ​

Manufacturers established outside the EU shall designate a single authorised representative (AR) within the EU.

10.12 — Person Responsible for Regulatory Compliance (PRRC) ​

Manufacturers shall ensure that they have at least one person responsible for regulatory compliance (PRRC) with the requisite expertise in the field of medical devices.

PRRC Responsibilities:

  • Ensure that the conformity of the devices is appropriately checked before release
  • Ensure that the technical documentation and the EU declaration of conformity are drawn up and kept up to date
  • Ensure that post-market surveillance obligations are fulfilled
  • Ensure that the reporting obligations for serious incidents and field safety corrective actions are fulfilled
  • Ensure that, in the context of investigations, the necessary support is provided to competent authorities

Official Sources ​