Skip to content

Periodic Safety Update Report (PSUR)

MDR Reference: Article 86; Annex III
BPG Rev.4 Reference: Section 7.2 (pp. 90-92)
Document Code: AIII-PSUR
Responsible Party: CRO
Service Mode: Write only

Applicability

  • Required for: Class IIa, IIb, and III devices
  • Update frequency (per Article 86):
    • Class IIa: at least every 2 years
    • Class IIb / Class III: at least annually
  • Class I devices: PMS Report per Article 85 (not PSUR)

Requirements Checklist

Content (per Article 86 and Annex III)

  • [ ] Summary of PMS data analysis for the reporting period
  • [ ] Conclusions from benefit-risk assessment based on PMS data
  • [ ] Main findings from PMCF activities
  • [ ] Volume of sales data (units sold) and estimation of usage population
  • [ ] Summary of complaints (categorised by type and frequency)
  • [ ] Serious incidents and FSCAs reported during period
  • [ ] Trend analysis against PMS Plan indicators and thresholds
  • [ ] Actions taken as result of PMS findings (CAPAs, design changes, labelling updates)
  • [ ] Conclusions and recommendations:
    • [ ] Risk-benefit ratio remains acceptable
    • [ ] Actions to optimise safety and performance
    • [ ] Any necessary updates to TD (CER, RMR, IFU, GSPR, etc.)
  • [ ] Reference to PMCF evaluation report findings
  • [ ] Updates to SSCP triggered (if applicable)

Lifecycle Integration

  • [ ] PSUR conclusions reflected in next CER update
  • [ ] PSUR findings feed back into risk management file
  • [ ] PSUR triggers PMS Plan updates where needed
  • [ ] Alignment with SSCP annual review
  • [ ] Alignment with NB surveillance schedule

Submission

  • [ ] Uploaded to EUDAMED (when functional)
  • [ ] Available to NB for assessment
  • [ ] In absence of EUDAMED: follow NB-specific submission arrangements (per MDCG 2021-1)

Key Regulatory References

  • MDR (EU) 2017/745, Article 86 (PSUR)
  • MDR (EU) 2017/745, Annex III (PMS documentation)
  • MDR (EU) 2017/745, Article 87 (Vigilance reporting)
  • MDCG 2022-21: Guidance on PSUR (including end of obligation to update)
  • MDCG 2021-1: EUDAMED interim arrangements
  • Team-NB BPG for TD Submission, Rev.4 (2026-04-21), Section 7.2

Common Pitfalls (from BPG Rev.4)

  • PSUR update frequency not per MDR Article 86 requirements
  • Incomplete trend analysis or missing connection to PMS Plan indicators
  • Benefits section is a re-statement of CER rather than analysis of new post-market data
  • Actions taken not clearly linked to specific PMS findings
  • Serious incident analysis lacking root cause investigation
  • PSUR not aligned with PMCF report and SSCP updates

Dependency

PSUR is a Phase 5 (post-certification lifecycle) deliverable:

  • Depends on PMS Plan (defines data collection scope)
  • Incorporates PMCF Report findings
  • Triggers updates to CER, RMR, IFU, SSCP
  • Feeds into NB annual surveillance review