NMPA Renewal Registration Dossier
Source: Notice 2021-121 Attachment 6.
Chinese: 延续注册.
Prerequisite: no certificate-listed changes (or changes controlled under QMS without altering listed particulars). Substantive changes require change registration/filing first.
Lean structure
- Regulatory information (prior certificate pack, declarations of no change, standards compliance notes)
- Non-clinical — updated PTR only if PTR changed during validity; leftover non-clinical commitments
- Clinical — only leftover clinical commitments on the certificate
Imported devices: home-country marketing authorization is not required for renewal.