Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices : Guidance for Industry and Food and Drug Administration Staff
Published: 2016-04-20
Status: Final Type: Guidance Document Category: Premarket (510(k) / PMA / De Novo / IDE) Topics: Premarket, 510(k), Labeling, Laboratory Tests, IVDs (In Vitro Diagnostic Devices) Docket: FDA-2015-D-0230
Official Source
https://www.fda.gov/regulatory-information/search-fda-guidance-documents/technical-performance-assessment-digital-pathology-whole-slide-imaging-devices PDF: https://www.fda.gov/media/90791/download
Official Full Text
Technical Performance Assessment of Digital Pathology Whole Slide Imaging Devices
This guidance represents the current thinking of the Food and Drug Administration (FDA or Agency) on this topic. It does not establish any rights for any person and is not binding on FDA or the public. You can use an alternative approach if it satisfies the requirements of the applicable statutes and regulations. To discuss an alternative approach, contact the FDA staff or Office responsible for this guidance as listed on the title page. 14
I. Introduction 16
FDA is issuing this guidance to provide industry and agency staff with recommendations regarding the technical performance assessment data that should be provided for regulatory evaluation of a digital whole slide imaging (WSI) system. This document does not cover the clinical submission data that may be necessary to support approval or clearance. This document provides our suggestions on how to best characterize the technical aspects that are relevant to WSI performance for their intended use and determine any possible limitations that might affect their safety and effectiveness. 24 Recent technological advances in digital microscopy, in particular the development of whole slide scanning systems, have accelerated the adoption of digital imaging in pathology, similar to the digital transformation that radiology departments have experienced over the last decade. FDA regulates WSI system manufacturers to help ensure that the images intended for clinical uses are reasonably safe and effective for such purposes. Essential to the regulation of these systems is the understanding of the technical performance of the WSI system and the components in the imaging chain, from image acquisition to image display and their effect on pathologist’s diagnostic performance and workflow. Prior to performing non-technical analytical studies (i.e., those using clinical samples) and clinical studies to evaluate a digital imaging system’s performance, the manufacturer should first determine the technical characteristics that are relevant to such performance for its intended use and determine any possible limitations that might affect its safety and effectiveness. This guidance provides recommendations 37 for the assessment of technical characteristics of a WSI device. 39 FDA's guidance documents, including this guidance, do not establish legally enforceable responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory requirements are cited. The use of the word should in Agency guidance means that something is suggested or recommended, but not required. 45
II. Background 47
For over a hundred years, the reference method for the diagnosis of cancer and many other critical clinical conditions has been histopathological examination of tissues using conventional light microscopy. This process is known as surgical pathology in the United States. 52 In surgical pathology, patient tissue from surgery, biopsy or autopsy goes through a process that includes dissection, fixation, embedding, and cutting of tissue into very thin slices which are then stained, for example by the hematoxylin and eosin (H&E) protocol, and permanently mounted onto glass slides. The slides are examined by a pathologist under a light microscope by dynamically adjusting the focus and using different magnifications. By integrating their interpretations obtained by microscopic examination of the tissue from all slides pertaining to a case, pathologists arrive at a diagnosis of the case. 61 WSI refers to the digitization of the stained entire tissue specimen on a glass slide. The glass slide is still prepared and stained just as for conventional light microscopy. Depending on the system used, various magnifications, scanning methodologies, hardware, and software are employed to convert the optical image of the slide into a digital whole slide image. With WSI, the pathologist views the image on a computer monitor rather than through the microscope oculars. 68
III. Scope 70
This document provides guidance regarding only the technical performance assessment of WSI systems for regulatory evaluation. WSI systems are defined here as those consisting of (a) an image acquisition subsystem that converts the content of a glass slide into a digital image file, and (b) a workstation environment for viewing the digital images. If not otherwise specified, the term “image” in the context of whole slide imaging refers to a pyramid structure consisting of multiple images at different resolutions. The baseline image has the highest resolution. This guidance is applicable for surgical pathology tasks performed in the anatomic pathology laboratory. It is intended to provide recommendations to industry and FDA staff regarding only the technical performance assessment data needed for the regulatory evaluation of a WSI device. This document is not meant to provide guidance for special stain techniques or fluorescence imaging or for the non-technical analytical studies (utilizing clinical 82 samples) or pivotal clinical studies necessary to support safety and effectiveness, nor does this guidance alone suffice to demonstrate safety and effectiveness of WSI systems. Interpretation of WSI images on mobile platforms is beyond the scope of this guidance. 86
IV. Policy 88
The following subsections of this section describe the technical performance assessment data FDA believes will facilitate the regulatory evaluation of a WSI device. 91 IV(A). Description and Test Methods for Each Component 93 This subsection details the descriptions and the test methods at the component level that should be included in the technical performance assessment of a WSI device. For purposes of this guidance only, a component is a piece of hardware, software, or a combination of hardware and software that processes the image signals flowing through the imaging chain. The concept of a component is based on the transformation of the image signals. For example, the digital imaging sensor is a hardware device that converts optical signals into digital signals. The image composition component is a software program that stitches sub-images together to form a whole slide image. A component and a physical device need not be in close physical proximity. For example, the light source component and the image optics component are usually tightly coupled within the same device, while the display calibration data is often distributed in both the color profile in the computer environment component and the on-screen display settings in the display component. 107 The components in a WSI device can be grouped in two subsystems: image acquisition and image display. The image acquisition subsystem digitizes the tissue slide as a digital image file. The image display subsystem converts the digital image file into optical signals for the human reader. In the paradigm of telemedicine, the digital image file can be electronically sent to a remote site for reading, so the image acquisition subsystem and the image display subsystem do not need to be physically coupled. Methods for independently testing the image acquisition and display subsystems are described in Section IV(B). 116 Sponsors should provide a block diagram of the components found in the WSI system in the premarket submission. A chart indicating the relationship among the components and the test methods utilized for the specific system characterization should also be provided. Diagram 1 on the following page is offered as an example block diagram of typical components found in current WSI systems. The components of a particular WSI system might not include all of those listed in the diagram or may include additional components. Sponsors are encouraged to provide additional diagrams, illustrations, and photographs of their devices as part of their submissions. 125 Diagram 1: Example block diagram of typical components found in current WSI 127 systems 129 131 133 135 137 139 141 143 145 147 149 151 153 155 157 159 161 163 165 167
IV(A)(1). Slide Feeder 168 IV(A)(1)(a). Description 171 The slide feeder is the mechanism(s) used to introduce the slide(s) to the scanner. For the slide feeder, sponsors should provide the following information, if applicable: · Configuration of the slide feed mechanism (a physical description of the equipment) o Slide configuration (physical description of the slide (i.e., custom or commercial off-the-shelf)) o Number of slides in queue (carrier) o Class of automation (e.g., robotics, pneumatics, etc.)
- User interaction o Hardware (e.g., loading of slides into carrier) o Software (e.g., does the system recognize the number of slides or is this specified by the user) o Feedback (e.g., alarms, notifications, etc.) o Failure Mode and Effects Analysis (FMEA) (including severity, likelihood, mitigations, etc.) 187 IV(A)(2). Light Source 189 IV(A)(2)(a). Description 191 The light source, including the light guide, generates and delivers light to the slide being imaged. The two major components are the lamp and condenser. For the light source, sponsors should provide the following information and specifications, if applicable: · Lamp o Bulb type (e.g., halogen, xenon arc, LED) o Manufacturer and model o Wattage o Spectral power distribution o Expected lifetime o Output adjustment control (electrical/electronic/mechanical) o Optical filter(s) § Type (e.g., heat blocking, polarization, neutral density, diffusing) o Manufacturer and model o Expected intensity variation (coefficient of variation ) § Over the duration of scanning a single slide § Over the course of a single workday § Over the lifetime of the device o Expected spectral variation § Over the duration of scanning a single slide § Over the course of a single workday § Over the lifetime of the device o Capability of tracking intensity and spectral degradation with lifetime · Condenser 214 o Illumination format (e.g., Kohler, critical) o Manufacturer and model o Numerical aperture o Focal length o Working distance 220 IV(A)(2)(b). Test Method 222 The following steps should be used to measure the spectral distribution of light incident on the slide. Position the input of a calibrated spectrometer or monochromator at the plane where the slide would be placed, centered on the illumination spot from the condenser. If desired, the light can be coupled into the spectrometer via light guide (e.g., fiber optic cable) or an integrating sphere. The measurement aperture should be at least as large as the anticipated field of view on the slide at the lowest magnification of the imaging optics. The wavelength accuracy and relative spectral efficiency of the spectrometer or monochromator in the wavelength range of 360-830 nm should be calibrated prior to measurements and reported. Plots of the measured spectrum with at least 10 nm spectral resolution should be provided, using radiometric units (e.g., spectral irradiance in W/cm2/nm, spectral radiance in W/sr/cm2/nm). 234 IV(A)(3). Imaging Optics 236 IV(A)(3)(a). Description 238 The imaging optics comprises the microscope objective and auxiliary lens(es) (e.g., tube lens), which optically transmit an image of the tissue from the slide to the digital image sensor. Sponsors should provide the following information and specifications, if applicable: · Optical schematic with all optical elements identified from slide (object plane) to digital image sensor (image plane)
- Microscope objective o Manufacturer o Type o Magnification o Numerical aperture (NA) o Focal length o Working distance · Auxiliary lens(es) o Manufacturer o Lens type o Focal length · Magnification of imaging optics: ISO 8039:2014 Optics and optical instruments — Microscopes — Magnification 258 IV(A)(3)(b). Test Methods 260 Sponsors should conduct the following tests in conformance with the International Standards, if applicable: · Relative irradiance of imaging optics at image plane per ISO 13653:1996 Optics and optical instruments – General optical test methods - Measurement of relative irradiance in the image field · Distortion per ISO 9039:2008 Optics and photonics — Quality evaluation of optical systems —Determination of distortion · Chromatic aberrations per ISO 15795:2002 Optics and optical instruments — Quality evaluation of optical systems — Assessing the image quality degradation due to chromatic aberrations 272 IV(A)(4). Mechanical Scanner Movement 274 IV(A)(4)(a). Description 276 The mechanical scanner addresses the physical characteristics of the stage upon which the glass slide is affixed. The key components include stage configuration, movement, and control. This information is relevant whether it is only the stage that is moving and the optics are stationary, or if there is movement on all axes. For the mechanical scanner, sponsors should provide the following information and specifications, if applicable: · Configuration of the stage (a physical description of the stage) o Stage size o Stage manufacturer and model number o Stage material (e.g., anodized aluminum) o Single multi-axis or multiple stacked linear stages (manufacturer and model number) o Type of guides or ways (e.g., bearings) o Sample retention mechanism (slide holder)
- Method of movement of the stage (e.g., stepper motor, servomotor, piezomotor, etc., coupled with belt, ball-screw, lead-screw, etc.) o Movement resolution for XY-axes o Movement in Z-axis o Speed range o Travel distance o Maximum scanning area o Localization and reading of bar code labels · Control of movement of the stage o Open or closed loop operation o Positional accuracy (calibration) and repeatability § Lost motion compensation (e.g., backlash) o Physical control (e.g., joystick) for single-slide, non-batch mode o Selection of area to be scanned (in accordance to image composition software) § whole slide § automatically determined area with tissue content 306
- Failure Mode and Effects Analysis (FMEA) (including severity, likelihood, mitigations, etc.) 309 IV(A)(4)(b). Test Method 311 Sponsors should demonstrate the mechanical performance of the stage with respect to positional repeatability and accuracy on all relevant axes, in accordance with ISO 230- 2:2014 Test code for machine tools—Part 2: Determination of accuracy and repeatability of positioning numerically controlled axes. 316 IV(A)(5). Digital Imaging Sensor 318 IV(A)(5)(a). Description 320 The digital image sensor is an array of photosensitive elements (pixels) that convert the optical signals of the slide to digital signals, which consist of a set of values corresponding to the brightness and color at each point in the optical image. Please provide the following information and specifications: · Sensor type (e.g., CMOS, CCD) and manufacturer · Pixel information/specifications o Number and dimensions of pixels o Design of color filter array § Configuration of color filter array § Spectral transmittance of color filter mask · Responsivity specifications o Relative response versus wavelength o Linearity o Spatial uniformity · Noise specifications o Dark current level (electrons per second) o Read noise (electrons)
- Readout rate (e.g., pixels per second, frames per second)
- Digital output format (e.g., bits per pixel, bits per color channel) 340 IV(A)(5)(b). Test Methods 342 Sponsors should conduct the following tests in conformance with the corresponding International Standards, if applicable: 345
- Opto-electronic conversion function per ISO 14524:2009 Photography — Electronic still-picture cameras — Methods for measuring optoelectronic conversion functions (OECFs)
- Noise measurements per ISO 15739:2013 Photography — Electronic still-picture imaging — Noise measurements 351
IV(A)(6). Image Processing Software 352 IV(A)(6)(a). Description 355 Image processing software refers to the embedded software components of the image acquisition device. It typically includes control algorithms for image capture and processing algorithms for raw data conversion into the digital image file. Sponsors should provide the following information and specifications, if applicable: · Exposure control · White balance · Color correction · Sub-sampling · Pixel-offset correction · Pixel-gain or flat-field correction · Pixel-defect correction 367 IV(A)(6)(b). Resources 369 See the guidance entitled “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices” (http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocument s/ucm089543.htm) for the information that should be provided. 374 IV(A)(7). Image Composition 376 IV(A)(7)(a). Description 378 Image composition is a step present in systems that produce whole slide images as opposed to individual fields of view. Whole slide scanning is typically performed in accordance with the positioning of a stage that moves in submicron steps. At each location of the stage movement, an image of the field of view is acquired. Images can be acquired with a degree of overlapping (redundancy) between them to avoid gaps in data collection. Images can also be acquired at different depths of focus followed by the application of focusing algorithms. At the end of this process, all acquired images are combined (stitched) together to create a composite high resolution image. There are a number of features that can affect this process, and they are listed below. Sponsors should provide a description of these features, if applicable: · Scanning method o Single objective or multiple miniature objectives in an array pattern o Scanning pattern: square matrix acquisition (tiling), line scanning, etc. o Overlap between scanned regions o Merging algorithms that stitch the aligned images together into a composite image file. Such algorithms may employ functions to align adjacent fields of view in accordance to the scanning pattern, overlap, etc. o Automatic background correction functions to eliminate the effect of non- 396 uniformities in the microscope’s illumination and image merging procedure. These non-uniformities if not corrected might create visible borders (seams and stitch lines) between the adjacent fields of view.
- Scanning speed: time to scan the whole slide. This time is dependent on selected magnification, and the amount of tissue on the glass slide. 401
- Number of planes at the Z-axis to be digitized (stack depth) 402 403 IV(A)(7)(b). Test Methods 404 405 Testing for image composition can be performed on a system level using special 406 calibration slides (such as grid patterns) that can test for line uniformity and focus 407 quality. Sponsors should provide the following outputs for these tests, if applicable: 408
- Images of digitized calibration slides 409
- Analysis of focus quality metrics 410
- Analysis of coverage of the image acquisition for the entire tissue slide 411 412 IV(A)(8). Image Files Formats 413 414 IV(A)(8)(a). Description 415 416 The final result from image acquisition can be a whole slide image consisting of a stack 417 of all acquired fields of view and magnifications during WSI. The complete digitized 418 image file usually occupies between 1-20 gigabytes of storage space depending on the 419 sample and the magnification of the objective lens used. Images can then be stored in a 420 number of ways and formats. Sponsors should provide the following information: 421
- Compression method (e.g., the wavelet-based JPEG2000 compression standard or 422 TIFF) 423
- Compression ratio: ratio of uncompressed to compressed file size. This metric 424 should be provided along with descriptive information on the data it was 425 measured from, since compression ratio is dependent on the content of the data 426 applied to. 427
- Compression type: lossless or lossy compression 428
- File format: can be formats easily accessible with public domain software such as 429 JPEG or TIFF, or can be proprietary formats only accessible with specific vendor 430 viewers. The file format depends on the file organization and related use. 431
- For systems that interact with DICOM-compliant software and hardware, 432 sponsors should provide a DICOM compatibility report. 433
- File organization: 434 o Single file with multi-resolution information (pyramidal organization) 435 o Stack of files at different magnifications 436 437 438 439
IV(A)(9). Image Review Manipulation Software 440 441 IV(A)(9)(a). Description 442 443 For the image review manipulation software, sponsors should provide the following 444 information, describing software features, if applicable. 445
- Continuous panning (moving in x-y space) and pre-fetching (buffering adjacent 446 images to speed up panning time) 447
- Continuous zooming (magnification) 448
- Discrete Z-axis displacement 449
- Ability to compare multiple slides simultaneously on multiple windows 450
- Ability to perform annotations 451
- Image enhancement such as sharpening functions 452
- Color manipulation, including color profile, white balance, color histogram 453 manipulation, and color filters 454
- Annotation tools 455
- Tracking of visited areas and annotations 456
- Digital bookmarks (revisit selected regions of interest) 457
- Virtual “multihead microscope” (this is when multiple pathologists 458 simultaneously review the same areas remotely) 459 460 IV(A)(9)(b). Resources 461 462 See the guidance entitled “Guidance for the Content of Premarket Submissions for 463 Software Contained in Medical Devices” 464 (http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocument 465 s/ucm089543.htm) for additional information on this subject. 466 467 IV(A)(10). Computer Environment 468 469 IV(A)(10)(a). Description 470 471 Computer environment refers to the workstation, including both hardware and software 472 components, that retrieves the digital image file and drives the display for the user to 473 review the images. Sponsors should provide the following information and 474 specifications, if applicable: 475
- Computer hardware 476
- Operating system 477
- Graphics card 478
- Graphics card driver 479
- Color management settings 480
- Color profile 481
- Display interface (e.g., DVI or DisplayPort) 482 483
IV(A)(11). Display 484 485 IV(A)(11)(a). Description 486 487 The final stage of a WSI system is the display component that presents the scanned image 488 to the pathologists for reading. Technically, display refers to the optoelectronic device 489 that converts the digital image signals in the RGB space into optical image signals. For 490 the display, sponsors should provide the following information and specifications, if 491 applicable: 492
Technological characteristics of the display device (e.g., in-plane switching LCD 493 panel with TFT active-matrix array with fluorescent backlight) 494
Physical size of the viewable area and aspect ratio 495
For transmissive displays, backlight type and properties including temporal, 496 spatial, and spectral characteristics 497
Frame rate and refresh rate 498
Pixel array, pitch, pixel aperture ratio and subpixel matrix scheme (e.g., chevron, 499 RGBW) 500
Subpixel driving to improve grayscale resolution (e.g., spatial and temporal 501 dithering) 502
Supported color spaces 503
Display Interface 504
User controls of brightness, contrast, gamma, color space, power-saving options, 505 etc. via the on-screen display (OSD) menu 506
Ambient light adaptation including the ambient light sensing method, 507 instrumentation, and software tool description 508
Touch screen technology including method, functionality, and any calibration or 509 periodical re-tuning requirements 510
Color calibration tools (sensor hardware and associated software), color profile, 511 and method for color management 512
Frequency and nature of quality-control tests to be performed by the user and/or 513 the physicist with associated action limits. 514 515 IV(A)(11)(b). Test Methods 516 517
User controls: Modes and settings of the display undergoing testing should be 518 specified, including brightness, contrast, gamma, white point, color space, etc. 519 See 2.1 Modified-Performance Modes, IDMS 1.03. 520
Spatial resolution: Measurements of the transfer of information from the image 521 data to the luminance fields at different spatial frequencies of interest typically 522 done by reporting the modulation transfer function. Non-isotropic resolution 523 properties should be characterized properly by providing two-dimensional 524 measurements or measurements along at least two representative axes. See 7.7 525 Effective Resolution, IDMS 1.03. 526
Pixel defects (count and map): Measurements (counts) and location of pixel 527 defects. This is typically provided as a tolerance limit. Pixel defects can interfere 528 with the visibility of small details in medical images. See 7.6 Defective Pixels, 529 IDMS 1.03. 530
Artifacts: Evaluate for image artifacts such as ghosting and/or image sticking 531 from displaying a fixed test pattern for a period of time. See 4.6 Artifacts and 532 Irregularities, IDMS 1.03. 533
Temporal response: Measurements of the temporal behavior of the display in 534 responding to changes in image values from frame to frame. Since these 535 transitions are typically not symmetric, rise and fall time constants are needed to 536 characterize the system. See 10.2.3 Gray-to-Gray Response Time, IDMS 1.03. 537
Maximum and minimum luminance (achievable and recommended): 538 Measurements of the maximum and minimum luminance that the device outputs 539 as used in the application under recommended conditions and the achievable 540 values if the device is set to expand the range to the limit. See 2.4 Vantage-Point 541 Suite of Measurement, IDMS 1.03. 542
Grayscale: Measurements of the mapping between image values and the 543 luminance. See 6.1 Grayscale, IDMS 1.03. 544
Luminance uniformity and Mura test: Measurements of the uniformity of the 545 luminance across the display screen. See 8.1.2 Sampled Vantage-Point Uniformity 546 and 8.2.3 Mura Analysis, IDMS 1.03. 547
Stability of luminance and chromaticity response with temperature and lifetime 548
Bidirectional reflection distribution function: Measurements of the reflection 549 coefficients of the display device. Specular and diffuse reflection coefficients can 550 be used as surrogates for the full bidirectional reflection distribution function. See 551 11.12 Diagnostic: Characterizing Hemisphere Uniformity, IDMS 1.03. 552
Gray Tracking: Chromaticity at different luminance levels as indicated by the 553 color coordinates in an appropriate units system (e.g., CIE u’v’). See AAPM Task 554 Group 196 Report. 555
Color scale: Color coordinates of primary and secondary colors as a function of 556 the digital driving level and their additivity. See 6. Gray- and Color-Scale 557 Measurement and 5.4 Color-Signal White, IDMS 1.03. 558
Color gamut volume: See 5.31 Volume-Color-Reproduction Capability, IDMS 559 1.03. 560 561 IV(A)(11)(c). Resources 562 563 Those interested in learning more about these types of display considerations should 564 consider reading: 565 566
IDMS 1.03 - Information Display Measurements Standard Version 1.03, 567 International Committee for Display Metrology, Society for Information Display, 568 www.icdm-sid.org 569 570
E. Samei, A. Badano, D. Chakraborty, K. Compton, C. Cornelius, K. Corrigan, 571
M. J. Flynn, B. Hemminger, N. Hangiandreou, J. Johnson, M. Moxley, W.
572
Pavlicek, H. Roehrig, L. Rutz, J. Shepard, R. Uzenoff, J. Wang, and C. Willis, 573 Assessment of display performance for medical imaging systems, Report of the 574 American Association of Physicists in Medicine (AAPM) Task Group 18, 575 Technical Report, AAPM (April 2005). 576 577
- IEC 62563-1:2009, Medical electrical equipment – Medical image display 578 systems – Part 1: Evaluation methods 579 580
- Amendment 1 to IEC 62563-1: Medical image display systems – Part 1: 581 Evaluation methods 582 583
- The guidance entitled “Guidance for Industry and FDA Staff: Display Accessories 584 for Full-Field Digital Mammography Systems-Premarket Notification (510(k)) 585 Submissions” 586 (http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceD 587 ocuments/ucm107549.htm). 588 589 IV(B). System-level Assessment 590 591 This subsection details the test methods at the system level that should be included in the 592 technical performance assessment of a WSI device. In this guidance, system refers to a 593 series of consecutive components in the imaging chain with clearly defined, measureable 594 input and output. For example, a system-level test can be designed for the image 595 acquisition subsystem, the image display subsystem, or a combination of both. The goal 596 of system-level tests is to assess the composite performance of a series of consecutive 597 components in the imaging chain. System-level tests should be conducted when the 598 component-level tests are either unfeasible or unable to capture the interplay between 599 components. 600 601 The common framework of the system-level tests described in this section is to compare 602 the system under test with an ideal system based on the same input, and then report the 603 difference between their outputs quantitatively. Designing such a system-level test 604 typically involves the following steps: (1) define the scope of the system and its input and 605 output, (2) define the input, which in most cases is a test target or phantom, (3) measure 606 the input to establish the ground truth that would be generated by an ideal system, (4) 607 measure the output of the system under test, and (5) calculate the errors between the truth 608 and the output with a quantitative metric. The framework of a typical system-level test is 609 shown in Diagram 2. Notice that the ideal system is a hypothetical device that generates 610 the perfect output with respect to the objective of the test such as color or focus. The 611 purpose of the ideal system is to define the intended behavior of the system under test. 612 The ideal system does not need to be implemented. Instead, the ideal system should be 613 simulated by a test method that establishes the truth of the input phantom. 614 615 616 617 618
Diagram 2: Framework of a typical system-level test. 619 620
621
622
623
624 625 IV(B)(1). Color Reproducibility 626 627 IV(B)(1)(a). Description 628 629 Color reproducibility is one of the key characteristics of a WSI system. The color 630 characteristics are determined by every component in the imaging chain. Therefore, the 631 color characteristics might be best evaluated at the system level. Color reproducibility 632 indicates the accuracy and precision of the color transformation from the tissue sample on 633 the slide to the image on the display. The colors of the tissue specimen should be 634 accurately and precisely reproduced on the display based on the color reproduction intent, 635 which should be clearly defined and justified by the sponsor. 636 637 IV(B)(1)(b). Test Methods 638 639 The WSI system should be tested with a target slide. The target slide should contain a set 640 of measurable and representative color patches. Ideally the color patches should have 641 similar spectral characteristics to stained tissue. The color patches should include a 642 grayscale ramp for evaluating the grayscale response. The truth of the color patches 643 should be measured with proper apparatuses separately. 644 645 For each color patch, the intended color (i.e., the expected output color based on the color 646 reproduction intent defined by the Sponsor) should be calculated based on the truth of the 647 color patches. 648 649 The target slide should be scanned and displayed by the WSI system. The output color of 650 each color patch should be measured from the display. 651 652 The three datasets – truth, intended color, and output color – should be compared and 653 analyzed. The sponsor should provide a rationale if the intended color is different from 654 the truth. 655 656 657 658 659 660 661 662 Error System under test Ideal system Input (Phantom) Output Truth f
Diagram 3: Framework of the system-level color reproducibility test. 663 664 665 666
667
668
669
670 671 672 IV(B)(1)(c). Resources 673 674 Useful references on the subject of color reproducibility can be found at the International 675 Color Consortium website http://www.color.org. 676 677 IV(B)(2). Spatial Resolution 678 679 IV(B)(2)(a). Description 680 681 Spatial resolution is another key characteristic of a WSI system. The goal of this system- 682 level test is to evaluate the composite optical performance of all components in the image 683 acquisition phase (i.e., from slide to digital image file). 684 685 IV(B)(2)(b). Test Methods 686 687 The following test is recommended for assessing spatial resolution of the image 688 acquisition phase: 689
- Resolution and spatial frequency response: ISO 12233:2014(E) — Photography 690 — Electronic still picture imaging — Resolution and spatial frequency responses. 691 692 IV(B)(3). Focusing Test 693
694
- The quality of focus in WSI can be affected by a number of inter-related factors, 695 including the scanning method and approaches for constructing a focus map. Due 696 to a trade-off between the number of focus points and the overall speed of the 697 scanning process, focusing is typically based on a sample of focus points, 698 determined automatically (auto-focus) or manually by the user. Since tissue can 699 have uneven depth, auto-focus algorithms are needed to detect and adjust for 700 different depths of focus. 701 702
- Data demonstrating that the focus quality is acceptable, even in the presence of 703 uneven tissue, should be provided. Such data with proper justification could be 704 derived from a phantom study, from clinical data, or both in a complementary 705 fashion. The technology of phantom construction for testing focus is under 706 development and this guidance will be updated as such technologies become 707 available. Sponsors could attempt to build their own phantoms for testing depth 708 Truth Meter Intended Color Output Color Accuracy Precision WSI under test Color Reproduction Intent Target Slide -
of focus for their device. Alternatively, sponsors could provide experimental data 709 using clinical tissue slides. Sampling of cases for such an experiment should be 710 enriched for uneven tissue cases within a range representative of typical 711 laboratory output. Alternative approaches for assessing the focus quality of a 712 WSI will be considered along with proper justification. In addition, the following 713 specifications should be provided, if applicable: 714 o Focus method: auto-focus for high-throughput or user-operated focus 715 points 716 o Instructions for the selection of manual focus points (if applicable), 717 including number of focus points and location in relation to a tissue 718 sample 719 o Metrics used to evaluate focusing and description of methods to extract 720 them 721 o Methods for constructing focus map from sample focus points 722 723 Diagram 4: Framework of the system-level focusing test. 724 725 726 727 728 729 730 731 732 733 IV(B)(4). Whole Slide Tissue Coverage 734 735 IV(B)(4)(a). Description 736 737 During the scan phase, WSI systems usually skip blank areas where tissue is absent in 738 order to reduce scan time and file size. The purpose of the whole slide tissue coverage 739 test is to demonstrate that all of the tissue specimen on the glass slide is included in the 740 digital image file. 741 742 IV(B)(4)(b). Test Method 743 744 Sponsors should include a test that demonstrates the completeness of the tissue coverage. 745 Sponsors should describe the test method and include the following items: 746
- Selection of the input tissue slide 747
- How to determine the complete coverage of the input tissue slide 748
- How to measure the actual coverage of the WSI output 749
- Calculate the ratio of the actual to complete coverage 750 751 752 753 754 Error WSI under test WSI with perfect focusing capability Phantom Slide Actual Focus
Optimal Focus f
Diagram 5: Framework of the system-level whole slide tissue coverage test 755 756
757
758
759
760
761
762 763 IV(B)(5). Stitching Error 764 765 IV(B)(5)(a). Description 766 767 Stitching is the technique that enables a WSI system to combine thousands of sub-images 768 into a single whole-slide image. Although during the scanning process a certain amount 769 of overlapping between adjacent sub-images is maintained for alignment purposes, 770 successful stitching relies on the texture present in the overlapped area. When the 771 stitching algorithm fails to align two sub-images seamlessly, the error may or may not be 772 perceivable by the human reader depending on whether noticeable stitching artifacts are 773 generated. Therefore, a system-level test should be conducted when assessing the 774 stitching quality of the WSI system. 775 776 IV(B)(5)(b). Test Methods 777 778 Sponsors should include a test that evaluates the stitching errors and include the 779 following items: 780
- Selection of the input test slide 781
- Method for sampling of the stitching boundaries where stitching errors might 782 occur 783
- How to determine the ideal stitching as the ground truth 784 o For example, the region of the stitching boundaries can be re-imaged in 785 one shot such that there is no stitching artifact. 786
- How to evaluate quality of the actual stitching based on the perfect stitching 787 o For example, compare the image of stitching boundaries with the perfect 788 one that does not have stitching artifact. The difference between these two 789 images can be used as a figure of merit of the stitching quality. 790 791 Diagram 6: Framework of the system-level stitching error test 792 793
794
795
796
797
798
799
800 Error WSI under test WSI with perfect stitching capability Tissue Slide Actual Stitching
Perfect Stitching f Error WSI under test WSI with complete coverage capability Tissue Slide Actual Coverage
Complete Coverage f 801 IV(B)(6). Turnaround Time 802 803 IV(B)(6)(a). Description 804 805 Turnaround time is the time required by the WSI system to execute a particular user 806 operation such as panning/zooming where the software and I/O (input/output) devices 807 retrieve image data, execute the computation, and refresh the image on the display. The 808 turnaround time starts when the user enters a command via a keyboard stroke or a mouse 809 click/movement and finishes when the image is completely updated on the display. 810 Turnaround time is important for a WSI system when fast and repetitive panning 811 operations are performed during a search task, which is delay-free in an optical 812 microscope. Prolonged, unpredictable turnaround time may impact the user’s diagnostic 813 performance. The user interface should properly prompt the user when the operation is 814 incomplete and the requested image is not available . The turnaround time may vary 815 greatly depending on the user-requested operation, image content, data size/location, 816 computer workload, display size, etc. The sponsor should report the typical turnaround 817 time as well as the test method and test conditions. 818
819 IV(C). User Interface 820 821 IV(C)(1). Description 822 823 The user interface covers all components and accessories of the WSI system with which 824 users interact while loading the slides and acquiring, manipulating, and reviewing the 825 images. It also includes preparing the system for use (e.g., unpacking, set up, 826 calibration), and performing maintenance. Elements of the user interface have been 827 noted in many of the preceding sections and include two broad categories: 828
- Options through which the user operates the WSI system, such as: 829 o Software menu options (e.g., scanning parameters) 830 o Physical controls (e.g., clips on the slide feeder) 831 o Connectors and connections (e.g., cables connecting system components) 832
- Information presented to the user through 833 o Visual displays (e.g., scanned image, software menus) 834 o Sounds (e.g., tone played when scanning completed) 835 o Instructions (e.g., software users’ manual) 836 o Labels 837 838 IV(C)(2). Test Methods 839 840 It is recommended that the analysis to identify the use-related hazards of the WSI system 841 include the consideration of use errors involving failure to acquire, perceive, read, 842 interpret, and act on information from the WSI system correctly or at all and the harm 843 that could be caused by such errors. A human factors/usability validation test should be 844 performed to demonstrate that representative users of the WSI system can perform 845 essential tasks and those critical to safety under simulated use conditions. 846
20
847 When selecting participants for validation testing, sponsors should carefully consider user 848 capabilities and expectations that could potentially impact the safe and effective use of 849 the WSI system. Examples of items that should be considered, if applicable, include 850 visual acuity and type of vision correction and the impact of expectations formed from 851 prior experience with other systems (e.g., optical microscope). 852 853 When selecting the critical tasks to be evaluated, sponsors should incorporate all known 854 use related errors and problems from similar devices (devices having similar 855 technological characteristics and indications for use) into the validation testing. 856 Consideration also should be given to whether task performance changes over time, and 857 if test duration needs to account for user fatigue. Examples might include a user altering 858 a task sequence in response to fatigue from repetitive image selection and manipulation 859 with mouse or keyboard. 860 861 When creating the simulated use conditions for validation testing, special consideration 862 should be given to the location of the WSI system primary workstation, its components, 863 their arrangement and how their locations affect user performance. Examples of location 864 considerations might include multiple monitors, a monitor with sub-optimal display 865 settings, or glare on a monitor from indoor lighting. 866 867 A human factors/usability validation test report should generally include the information 868 found in Table 1. 869 870 Table 1: Items a Human Factors/Usability Validation Test Report Should Include 871 872 Section Contents Intended device users, uses, use environments, and training - Intended user population(s) and critical differences in capabilities between multiple user populations · Intended uses and operational contexts of use · Use environments and key considerations - Training intended for users and provided to test participants Device user interface - Graphical depiction (drawing or photograph) of device user interface · Verbal description of device user interface Summary of known use problems · Known problems with previous models · Known problems with similar devices - Design modifications implemented in response to user difficulties User task selection, characterization and prioritization - Risk analysis methods - Use-related hazardous situation and risk summary - Critical tasks identified and included in HFE/UE validation tests Summary of formative evaluations · Evaluation methods · Key results and design modifications implemented - Key findings that informed the HFE/UE validation testing protocol Validation testing - Rationale for test type selected (i.e., simulated use or clinical evaluation)
- Number and type of test participants and rationale for how they represent the intended user populations - Test goals, critical tasks and use scenarios studied - Technique for capturing unanticipated use errors - Definition of performance failures - Test results: Number of device uses, success and failure occurrences - Subjective assessment by test participants of any critical task failures and difficulties - Description and analysis of all task failures, implications for additional risk mitigation Conclusion A statement to the effect that “The <device name/model> has been found to be reasonably safe and effective for the intended users, uses and use environments” should be included under the following conditions: - The methods and results described in the preceding sections support this conclusion.
- Any residual risk that remains after the validation testing would not be further reduced by modifications of design of the user interface (including any accessories and the Instructions for Use (IFU)), is not needed, and is outweighed by the benefits that may be derived from the device’s use. 873 Recommended methods for performing a human factors/usability validation test are 874 described in the resources listed in section IV(C)(3) entitled “Resources” directly below. 875 The goal of testing is to assure that users can operate the WSI system successfully for the 876 intended uses without negative clinical consequences to the patient and that potential use 877 errors or failures have been eliminated or reduced. 878 879 IV(C)(3). Resources 880 881 FDA recognizes standards published by national and international organizations that 882 apply human factors engineering/usability engineering (HFE/UE) principles to device 883 design and testing. The recognized standards listed below provide suggestions on 884 conducting an analysis of use-related hazards and a human factors/usability validation 885 test to assess the safety and effectiveness of the final device design. 886 887
- ISO 14971:2007, Medical Devices – Application of Risk Management to Medical 888 Devices: Provides systematic process to manage the risks associated with the use 889 of medical devices. 890
- AAMI/ANSI HE75:2009, Human Factors Engineering – Design of Medical 891 Devices: Comprehensive reference of recommended practices related to human 892 factors design principles for medical devices. 893
- IEC 62366-1:2015, Medical devices – Application of usability engineering to 894 medical devices: Describes the process to conduct medical device usability testing 895 and incorporate results into a risk management plan. 896 In addition, FDA has published guidance with human factors related recommendations to 897 assist manufacturers and facilitate premarket review. The guidance entitled “Guidance 898 for the Content of Premarket Submissions for Software Contained in Medical Devices” 899 (http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocument 900 s/ucm089543.htm). This guidance document provides recommendations to industry 901 regarding premarket submissions for software devices, including stand-alone software 902 applications and hardware-based devices that incorporate software. It includes test 903 methods to assure that the software conforms to the needs of the user and to check for 904 proper operation of the software in its actual or simulated use environment. 905 906 IV(D). Labeling 907 908 The premarket application must include labeling in sufficient detail to satisfy the 909 requirements of 21 CFR Part 801 and 21 CFR 809.10. The labeling includes 910 supplementary information necessary to use and care for the WSI system such as 911 instruction books or direction sheets and software user manuals. 912 913
Although instructions, labeling, and training can influence users to use devices safely and 914 effectively, they should not be the primary strategy used to control risk. Modification of 915 the user interface design is a more effective approach to mitigate use-related hazards. 916 917 IV(D)(1). Test Methods 918 919 It is recommended that studies on labeling and training be conducted separately from 920 other human factors/usability validation testing. Human factors/usability validation 921 testing should be conducted with the final version of the labeling and related materials. 922 Timing and content of training should be consistent with that expected of actual users. 923 924 IV(D)(2). Resources 925 926 FDA has published several guidance documents on labeling to facilitate premarket 927 review and assist manufacturers. 928
- The guidance entitled “Labeling - Regulatory Requirements for Medical Devices” 929 (http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/ 930 GuidanceDocuments/UCM095308.pdf). 931 o This publication covers labeling issues that device manufacturers, 932 reconditioners, repackers, and relabelers should consider when a product 933 requires labeling. Labeling includes adequate instructions for use, 934 servicing instructions, adequate warnings against uses that may be 935 dangerous to health, or information that may be necessary for the 936 protection of users. 937
- The guidance entitled “Device Labeling Guidance #G91-1 (blue book memo)” 938 (http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceD 939 ocuments/ucm081368.htm). 940 o This guidance is intended to ensure the adequacy of, and consistency in 941 device labeling information. It is intended for use by industry in preparing 942 device labeling. 943
- The guidance entitled “Human Factors Principles for Medical Device Labeling” 944 (http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/ 945 GuidanceDocuments/UCM095300.pdf). 946 o This report presents the principles of instruction, human factors, and 947 cognitive psychology that are involved in designing effective labeling for 948 medical devices. 949 950 IV(E). Quality Control 951 952 Sponsors should provide information on the quality control procedures, including 953 frequency and testing methods to be performed by the laboratory technologists and/or 954 field engineers with associated quantitative action limits. Discussions of tests for 955 constancy should include discussions of the slide feeder and scanning mechanisms, 956 coverage of the entire tissue slide, the bar code reader, the light source, the imaging 957 sensor device, and the calibrations at the component and system level. A detailed quality 958 control manual should be provided. 959